Commands listed in alphabetical order
Examples
DISplay:PLOTV
iew1:ZOOM:XAXIS:TO 2.5
sets the end of the zoom x-axis
to 2.5 V.
DISplay:PLOT
View3:ZOOM:XAXIS:TO?
might return
:DISPLAY:PLOTVIEW3:ZOOM:X AXIS:TO 2.2144883507508
indicating the end of the zoom x-axis is 2.21 V.
DISplay:PL
OTView<x>:ZOOM:YAXIS:FROM
This command sets or queries the bottom value of the zoom y-axis in the specied
plot view.
Group
Zoom
Syntax
DISplay:PLOTView<x>:ZOOM: YAXIS:FROM <NR3>
DISplay:PLOTView<x>:ZOOM: YAXIS:FROM?
Arguments
<NR3> is the bottom value of the zoom y-axis.
Examples
DISplay:PLOTView1:ZOOM:YA XIS:FROM -1.0E+20 sets the bottom value
of the zoom y-axis to -1E+20.
DISplay:PLOTView1:ZOOM:YA XIS:FROM? might return
:DISPLAY:PLOTVIEW1:ZOOM:Y AXIS:FROM -1.0E+21 indicating
the bottom value of the z oom y-axis is -1E+21.
DISplay:PLOTView<x>:ZOOM:YAXIS:TO
This command sets or queries the top value of the zoom y-axis in the specied
plot view.
Group
Zoom
Syntax
DISplay:PLOTView<x>:ZOOM: YAXIS:TO <NR3>
DISplay:PLOTView<x>:ZOOM: YAXIS:TO?
Arguments
<NR3> is the top value of the zoom y-axis.
Examples
DISplay:PLOTView1:ZOOM:YA XIS:TO 100 sets the top value of the zoom
y-axis to 100.
2-262 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Other manuals for Tektronix MSO54

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO54 and is the answer not in the manual?

Tektronix MSO54 Specifications

General IconGeneral
BrandTektronix
ModelMSO54
CategoryTest Equipment
LanguageEnglish

Related product manuals