Commands listed in alphabetical order
Syntax
POWer:POWer<x
>:NDUTYCYCLE:EDGEType
{RISE | FALL | BOTH}
POWer:POWer<x>:NDUTYCYCLE :EDGEType?
Examples
POWer:POWer1:NDUTYCYCLE:E DGEType RISE se ts the clock edge type as rise
for the negative duty cycle measurement.
POWer:POWer<x>:NDUTYCYCLE:INPUTSOurce
This command sets or queries the input source for negative duty cycle
measurement in the specified power m easurement badge. The power measurement
badge is specified by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:NDUTYCYCLE :INPUTSOurce {CH<x> | MATH<x> |
REF<x>}
POWer:POWer<x>:NDUTYCYCLE :INPUTSOurce?
Arguments
CH<x> = A channel specifier; <x> is 1 through 8 and is limited by the number
of FlexChannels in your instrument.
MATH<x> = A math waveform specifier; <x> is ≥1.
REF<x> = A reference waveform specifier; <x> is ≥1.
Examples
POWer:POWer1:NDUTYCYCLE:I NPUTSOurce CH3 sets the input source for
negative duty cycle measurement as CH3 for the power measurement badge
Power 1.
POWer:POWer<x>:NPULSEWIDTH:INPUTSOurce
This command se t s or queries the input source for negative pulse width
measurement in the specified power m easurement badge. The power measurement
badge is specified by x.
Conditions
Requires option 5-PWR, S
UP5-PWR, or 5-PS2.
Group
Power