Commands listed in alphabetical order
MEASUrement:M
EAS1:COMMONMode:SOURCEs?
might return
:MEASUREMENT:MEAS1:COMMONM ODE:SOURCES DOUBLE indicating sources
are set to double.
MEASUrement
:MEAS<x>:CYCLemode
This command sets or queries the cycle mode for the measurement. Measurements
are specified by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:CYCLem ode {RECORD|CYCLE}
MEASUrement:MEAS<x>:CYCLem ode?
Arguments
RECORD specifies that the measurement is taken o ver the whole record.
CYCLE specifies that measurements are taken on each cycle of the source.
Examples
MEASUrement:MEAS1:CYCLemod e CYCLE sets the measurement to be taken
over each cycle of the source.
MEASUrement:MEAS1:CYCLemod e? might return
:MEASUREMENT:MEAS1:CYCLEMO DE RECORD indicating the measurement is
taken over the whole record.
MEASUrement:MEAS<x>:DELay:EDGE<x>
This command sets or queries the 'to edge' type when EDGE<x> is EDGE1 and
the 'from edge' type when EDGE<x> is EDG2, for the measurement when the
measurement type is DELAY. Measurements are specified by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:DELay: EDGE<x>
{FALL|RISe|BOTH|SAMEas|OPP ositeas}
MEASUrement:MEAS<x>:DELay: EDGE<x>?
Related Commands
MEASUrement:MEAS<x>:TOEdge
MSO54, MSO56, MSO58, MSO58LP Programmer 2-441