Commands listed in alphabetical order
POWer:POWer<
x>:RESUlts:CURRentacq:FREQUENCY? (Query Only)
This command queries the fundamental frequency for the specified power
measurement badge. The power measurement badge is specified by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:RESUlts:CU RRentacq:FREQUENCY? "harmonics"
Examples
POWer:POWer1:RESUlt
s:CURRentacq:FREQUENCY? "harmonics"
might
return 100.0067656931537E+3, indicating the fundamental frequency for the
power measurement badge 1.
POWer:POWer<x>:RE
SUlts:CURRentacq:IRMS? (Query Only)
This command queries the RMS current value for the specified power
measurement badge. The power measurement badge is specified by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:RESUlts:CU RRentacq:IRMS? "harmonics"
Examples
POWer:POWer1:RESUlts:CU
RRentacq:IRMS? "harmonics"
might return
1.4149980733491, indicating the RMS current value for the power measurement
badge 1.
POWer:POWer<x>:RESUlts
:CURRentacq:MAXimum? (Query Only)
This command queries the maximum value of the current acquisition for the
measurement parameter in the specified power measurement badge. The power
measurement badge is sp
ecified by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
2-570 MSO54, MSO56, MSO58, MSO58LP Programmer