EasyManuals Logo
Home>Tektronix>Test Equipment>MSO54

Tektronix MSO54 Programmer's Manual

Tektronix MSO54
1026 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #586 background imageLoading...
Page #586 background image
Commands listed in alphabetical order
POWer:POWer<
x>:RESUlts:CURRentacq:FREQUENCY? (Query Only)
This command queries the fundamental frequency for the specied power
measurement badge. The power measurement badge is specied by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:RESUlts:CU RRentacq:FREQUENCY? "harmonics"
Examples
POWer:POWer1:RESUlt
s:CURRentacq:FREQUENCY? "harmonics"
might
return 100.0067656931537E+3, indicating the fundamental frequency for the
power measurement badge 1.
POWer:POWer<x>:RE
SUlts:CURRentacq:IRMS? (Query Only)
This command queries the RMS current value for the specied power
measurement badge. The power measurement badge is specied by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
Group
Power
Syntax
POWer:POWer<x>:RESUlts:CU RRentacq:IRMS? "harmonics"
Examples
POWer:POWer1:RESUlts:CU
RRentacq:IRMS? "harmonics"
might return
1.4149980733491, indicating the RMS current value for the power measurement
badge 1.
POWer:POWer<x>:RESUlts
:CURRentacq:MAXimum? (Query Only)
This command queries the maximum value of the current acquisition for the
measurement parameter in the specied power measurement badge. The power
measurement badge is sp
ecied by x.
Conditions
Requires option 5-PWR, SUP5-PWR, or 5-PS2.
2-570 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Other manuals for Tektronix MSO54

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO54 and is the answer not in the manual?

Tektronix MSO54 Specifications

General IconGeneral
BrandTektronix
ModelMSO54
CategoryTest Equipment
LanguageEnglish

Related product manuals