Page #298 background imageLoading...
Page #298 background image
Commands listed in alphabetical order
Syntax
DISplay:VARpe
rsist <NR 3>
DISplay:VARpersist?
Related Commands
DISplay:PERSistence
Arguments
<NR3> indic
ates the persistence decay time and ranges from 0.5 to 100.
Examples
DISPLAY:VARPERSIST 5 sets the persistence decay time to 5.
DISPLAY:VARPERSIST? might return :DISPLA Y:VARPERSIST 3.0000E-01,
indicating that persistence decay time is currently set to 0.300.
DISplay:WAVEView<x>:BUS:B<x>:STATE
Sets or queries the state of the specied bus in the specied waveview.
NOTE.
WAVEView<x>
is the specied waveview and must be
WAVEView1
.
Group
Display Control
Syntax
DISplay:WAVEView<x>:BUS:B <x>:STATE {OFF|ON|0|1}
DISplay:WAVEView<x>:BUS:B <x>:STATE?
Arguments
0 turns specied bus off.
1 turns the specied bus on.
ON turns the specied bus on.
OFF turns specied bus off.
Examples
DISplay:WAVEView1:BUS:B1: STATE OFF turns specied bus off.
DISplay:WAVEView1:BUS:B1: STATE? might return
:DISPLAY:WAVEVIEW1:BUS:B1 :STATE 1 indicating the specied bus is on.
DISplay:WAVEView<x>:BUS:B<x>:VERTical:POSition
Sets or queries t he vertical position of the specied bus in the specied waveview.
NOTE.
WAVEView<x>
is the specied waveview and must be
WAVEView1
.
2-282 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Other manuals for Tektronix MSO54

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO54 and is the answer not in the manual?

Tektronix MSO54 Specifications

General IconGeneral
BrandTektronix
ModelMSO54
CategoryTest Equipment
LanguageEnglish

Related product manuals