Commands listed in alphabetical order
Examples
MEASUREMENT:G
ATING:HYSTERESIS 40.0e -3
sets the hysteresis to 4 0 mV.
MEASUREMENT:GATING:HYSTERE SIS? might return
:MEASUREMENT
:GATING:HYSTERESIS 30.0000E-3
indicating the hysteresis
is 30 mV.
MEASUrement:GATing:LOGICSource
This comman
d sets or queries the gating data source used for logic gating.
Group
Measurement
Syntax
MEASUrement:GATing:LOGICSo urce {CH<x>|MATH<x>|REF<x >}
MEASUrement:GATing:LOGICSo urce?
Arguments
Arguments are the logic gating source.
Examples
MEASUREMENT:GATING:LOGICSO URCE MATH1 sets the logic source to MATH 1.
MEASUREMENT:GATING:LOGICSO URCE? might return
:MEASUREMENT:GATING:LOGICS OURCE CH6 indicating the logic source is
channel 6.
MEASUrement:GATing:MIDRef
This command sets or queries the global gating m id ref value used for logic gating.
Group
Measurement
Syntax
MEASUrement:GATing:MIDRef <NR3>
MEASUrement:GATing:MIDRef?
Arguments
<NR3> is the mid ref value for gating.
Examples
MEASUREMENT:GATING:MIDREF 2.0 set the gating mid ref to 2.0 V.
MEASUREMENT:GATING:MIDREF? might return
:MEASUREMENT:GATING:MIDREF 1.5000 indicating the gating mid ref is 1.5 V.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-407

Table of Contents

Other manuals for Tektronix MSO54

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO54 and is the answer not in the manual?

Tektronix MSO54 Specifications

General IconGeneral
BrandTektronix
ModelMSO54
CategoryTest Equipment
LanguageEnglish

Related product manuals