Commands listed in alphabetical order
BUS:B<x>:MIL
1553B:RESPonsetime:MAXimum
This command sets or queries the maximum response time to a valid command
issued for the spe cied MIL-STD-1553 bus. The bus is speci ed by x.
Conditions
Requires option SR-AERO.
Group
Bus
Syntax
BUS:B<x>:MIL1553B:RESPonse time:MAXimum <NR3>
BUS:B<x>:MIL1553B:RESPonse time:MAXimum?
Related Commands
BUS:B<x>:MIL1553B:RESPonsetime:MINimum
Arguments
<NR3> is a oating point number that species the maximum response time, in
seconds.
Examples
BUS:B1:MIL1553B:RESPONSETI ME:MAXIMUM 15.0E-6 species the
maximum response time to a valid command received to be 15.0 microseconds.
BUS:B1:MIL1553B:RESPONSETI ME:MAXIMUM? might return
:BUS:B1:MIL1553B:RESPONSET IME:MAXIMUM 12.000E-6, indicating a
maximum response time of 12 microseconds.
BUS:B<x>:MIL1553B:RESPonsetime:MINimum
This command sets or queries the minimum response time to a valid command
issued for the spe cied MIL-STD-1553 bus. The bus is speci ed by x.
Conditions
Requires option SR-AERO.
Group
Bus
Syntax
BUS:B<x>:MIL1553B:RESPonse time:MINimum <NR3>
BUS:B<x>:MIL1553B:RESPonse time:MINimum?
Related Commands
BUS:B<x>:MIL1553B:RESPonsetime:MAXimum
MSO54, MSO56, MSO58, MSO58LP Programmer 2-157

Table of Contents

Other manuals for Tektronix MSO54

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO54 and is the answer not in the manual?

Tektronix MSO54 Specifications

General IconGeneral
BrandTektronix
ModelMSO54
CategoryTest Equipment
LanguageEnglish

Related product manuals