Commands listed in alphabetical order
Examples
MEASUrement:M
EAS1:CLOCKRecovery:CONSTCLOC KMODe FIXED
specifies
that clock recovery uses the value set by the user as the clock frequency.
MEASUrement:
MEAS1:CLOCKRecovery:CONSTC LOCKMODe?
might return
:MEASUREMENT:MEAS1:CLOCKRE COVERY:CONSTCLOCKMODE MEA N indicating
that clock recovery uses the mean of the clock signal as the clock frequency.
MEASUremen
t:MEAS<x>:CLOCKRecovery:DAMPing
This command sets or queries the da mping value used when PLL clock recovery
is used for the measurement. Measuremen ts are specified by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:CLOCKR ecovery:DAMPing <NR3>
MEASUrement:MEAS<x>:CLOCKR ecovery:DAMPing?
Arguments
<NR3> is the clock recovery damping value.
Examples
MEASUrement:MEAS1:CLOCKRec overy:DAMPing 700.0E-3 sets the
damping value to 0.70.
MEASUrement:MEAS1:CLOCKRec overy:DAMPing? might return
:MEASUREMENT:MEAS1:CLOCKRE COVERY:DAMPING 700.0000E-3 indicating
thedampingvalueis0.70.
MEASUrement:MEAS<x>:CLOCKRecovery:DATAPath
This command sets or queries the file containing the data pattern used when
known data pattern clock recovery is used for the measurement. Measurements
are specified by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:CLOCKR ecovery:DATAPath <QString >
MEASUrement:MEAS<x>:CLOCKR ecovery:DATAPath?
Arguments
<QString> is the file containing the data pattern to be used for known data
pattern clock recovery.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-433