Commands listed in alphabetical order
DISplay:MATH
FFTView<x>:CURSor:BSOUrce? (Query Only)
This command queries the cursor source for plot cursor B.
Group
Cursor
Syntax
DISplay:MATHFFTView<x>:CU RSor:BSOUrce?
Examples
DISplay:MATHFFTView1:CURS or:BSOUrce? might return
:DISPLAY:MATHFFTVIEW1:CUR SOR:BSOURCE MATH1 indicating the source
is MATH 1.
DISplay:MATHFFTView<x>:CURSor:DDT? (Query Only)
This command queries the delta Y over delta X c ursor readout value of the
specied cursor in the specied view.
Group
Cursor
Syntax
DISplay:MATHFFTView<x>:CU RSor:DDT?
Examples
DISplay:MATHFFTView<x>:CU RSor:DDT? might return
:DISPLAY:MATHFFTVIEW1:CUR SOR:DDT 9.91E+37 indicating the
delta Y over delta X cursor readout value is 9.91E+37.
DISplay:MATHFFTView<x>:CURSor:FUNCtion
This command sets or queries the cursor type of the specied cursor in the
specied view.
Group
Cursor
Syntax
DISplay:MATHFFTView<x>:CU RSor:FUNCtion
{WAVEform|VBArs|HBArs|SCR EEN}
DISplay:MATHFFTView<x>:CU RSor:FUNCtion?
Arguments
WAVEFORM species to display the paired cursors in YT display format for
measuring waveform amplitude and time. Waveform cursors are attached to the
waveform.
2-234 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Other manuals for Tektronix MSO54

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO54 and is the answer not in the manual?

Tektronix MSO54 Specifications

General IconGeneral
BrandTektronix
ModelMSO54
CategoryTest Equipment
LanguageEnglish

Related product manuals