EasyManuals Logo
Home>Tektronix>Test Equipment>MSO54

Tektronix MSO54 Programmer's Manual

Tektronix MSO54
1026 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #878 background imageLoading...
Page #878 background image
Commands listed in alphabetical order
NOTEQual argum
ent sets the oscilloscope to trigger if a runt pulse is d etected with
width greater than or less than (but not equal to) the time period specied in
TRIGger:{A|B}:RUNT:WIDth within a ±5% tolerance.
Examples
TRIGger:A:RUNT:WHEn MORETHAN sets the runt trigger to occur when the
oscilloscope detects a runt in a pulse wider than the specied width.
TRIGger:A:RUNT:WHEn? might return :TRIGGER:A:PULSE:RUNT:WHEN
OCCURS
indicating that a runt trigger will occur if the oscilloscope detects a runt
of any detectable width.
TRIGger:{A|B}:RUNT:WIDth
This command species the width, in seconds, for a runt trigger.
Group
Trigger
Syntax
TRIGger:{A|B}:RUNT:WIDth <NR3>
TRIGger:{A|B}:RUNT:WIDth?
Related Commands
TRIGger:{A|B}:RUNT:WHEn
Arguments
<NR3> is a oating point number that species the minimum width, in seconds.
Examples
TRIGger:A:RUNT:WIDth 15E-6 sets the m inimum width o f the pulse runt
trigger to 15 µs.
TRIGger:A:RUNT:WIDth? might return :TRIGGER:A:PULSE:RUNT:WIDT H
2.0000E-09
indicating that the minimum width of a pulse runt trigger is 2 ns.
TRIGger:{A|B}:SETHold:CLOCk:EDGE
This command species the clock edge polarity for setup and hold triggering.
Group
Trigger
Syntax
TRIGger:{A|B}:SETHold:CLO Ck:EDGE {FALL|RISe}
TRIGger:{A|B}:SETHold:CLO Ck:EDGE?
2-862 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Other manuals for Tektronix MSO54

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO54 and is the answer not in the manual?

Tektronix MSO54 Specifications

General IconGeneral
BrandTektronix
ModelMSO54
CategoryTest Equipment
LanguageEnglish

Related product manuals