EasyManuals Logo
Home>Tektronix>Test Equipment>MSO54

Tektronix MSO54 Programmer's Manual

Tektronix MSO54
1026 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #795 background imageLoading...
Page #795 background image
Commands listed in alphabetical order
TRIGger:{A|B
}:BUS:B<x>:ETHERnet:CONDition
This command species a eld or condition within an Ethernet frame to trigger
on. The bus number is specied by <x>.
Conditions
Requires the SR-ENET Triggering and Analysis application.
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:ETH ERnet:CONDition
{SFD|MACADDRess|MACLENgth| IPHeader|TCPHeader|DATa|EOP|
IDLe|FCSError|QTAG}
TRIGger:{A|B}:BUS:B<x>:ETH ERnet:CONDition?
Related Commands
Most of the other TRIGger: A:BUS:B<x>:ETHER net
commands are impacted
by the setting of this command.
Arguments
SFD Start of frame delimiter.
MACADDRess MAC addresses eld.
MACLENgth —MAC
length/type eld.
IPHeader —IPheadereld. This argument is only available when PROTOCOL
is set to IPv4.
TCPHeader TCP header eld. This argument is only available when
PROTOCOL is s
et to IPv4.
DATa TCP/IPv4 or MAC protocol client data eld. If the protocol is set to
OTHER, then
DATa refers to the M AC client data.
EOP End of Packet eld.
IDLe —Idleeld.
FCSError Frame Check Sequence Error (CRC) eld.
QTAG IEEE 802.1Q (VLAN) control information eld. In order to use QTAG as
a trigger condition, the frame type must be set to QTAG ).
Examples
TRIGger:A:BUS:B1:ETHERnet: CONDition MACADDRess sets the trigger
eld to
MACADDRess.
TRIGger:A:BUS:B1:ETHERnet: CONDition? might return DATA.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-779

Table of Contents

Other manuals for Tektronix MSO54

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO54 and is the answer not in the manual?

Tektronix MSO54 Specifications

General IconGeneral
BrandTektronix
ModelMSO54
CategoryTest Equipment
LanguageEnglish

Related product manuals