Commands listed in alphabetical order
TRIGger:A:LOG
Ic:WHEn?
might return :TRIGGER:A:LOGIC:WHEN TRUE
indicating a trigger when the logic is true.
TRIGger:{A|B}:LOGICPattern:{CH<x>|CH<x>_D<x>}
This command sets or queries the Logic Pattern that is used along with the Dene
Logic choice (
LOGIc:FUNCtion) to determine when the logic trigger occurs.
Group
Trigger
Syntax
TRIGger:{A|B}:LOGICPattern :{CH<x>|CH<x>_D<x>} {HIGH |LOW|X}
TRIGger:{A|B}:LOGICPattern :{CH<x>|CH<x>_D<x>}?
Arguments
HIGH species the logic high.
LOW speciesthelogiclow.
X speciesadon'tcarestate.
Examples
TRIGger:A:LOGICPattern:CH1 HIGH species triggering on a logic high.
TRIGger:A:LOGICPattern:CH1 ? might return
:TRIGGER:A:LOGICPATTERN:CH 1 X indicating a don’t care state for channel 1.
TRIGger:A:LOGICPattern:CH1 _D0 HIGH species a logic high.
TRIGger:A:LOGICPattern:CH1 _D0? might return
:TRIGGER:A:LOGICPATTERN:CH 1_D0 X indicating a don’t care.
TRIGger:{A|B}:LOWerthreshold:CH<x>
This command sets or queries the A or B lower trigger level threshold for the
channel, specied by x.
Group
Trigger
Syntax
TRIGger:{A|B}:LOWerthresho ld:CH<x> <NR3>
TRIGger:{A|B}:LOWerthresho ld:CH<x>?
Related Commands
TRIGger:{A|B}:UPPerthreshold:CH<x>
Arguments
<NR3> species the threshold voltage in user units.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-855

Table of Contents

Other manuals for Tektronix MSO54

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO54 and is the answer not in the manual?

Tektronix MSO54 Specifications

General IconGeneral
Analog Channels4
Sample Rate6.25 GS/s
Vertical Resolution12-bit
Form FactorBenchtop
Probe InterfaceTekVPI
Operating Temperature0 °C to +50 °C
Bandwidth1 GHz
Record Length62.5 Mpoints
Display15.6 inch HD
Waveform Update Rate>500, 000 wfms/s
Trigger TypesEdge, Pulse, Runt, Logic, Serial
Serial Bus SupportI2C, SPI, RS-232/422/485/UART, CAN/CAN FD/LIN, USB 2.0, Ethernet
ConnectivityUSB, Ethernet
Input Impedance1 MΩ
Power Requirements100-240 VAC, 50/60 Hz
Maximum Input Voltage400 V (DC + peak AC)

Related product manuals