Commands listed in alphabetical order
SEARCH:SEARC
H<x>:TRIGger:A:BUS:CAN:DATa:VALue
This command sets or queries the binary data value to be used when searching on
a CAN bus signal. The search condition must be set to IDANDDATA OR DATA.
Group
Search and Mark
Syntax
SEARCH:SEARCH<x>:TRIGger:A :BUS:CAN:DATa:VALue <QString>
Arguments
<QString>
Examples
SEARCH:SEARCH1:TRIGger:A:B US:CAN:DATa:VALue "1111" sets the data
value to 1111.
SEARCH:SEARCH1:TRIGger:A:B US:CAN:DATa:VALue? might return
:SEARCH:SEARCH1:TRIGGER:A: BUS:CAN:DATA:VALUE "1010" indicating
the data value is 1010.
SEARCH:SEARCH<x>:TRIGger:A:BUS:CAN:ERRType
This command sets or queries the type of error condition for a CAN bus to
search on. The search number is specied by x. The search condition must be
set to ERRor.
Conditions
Requires the SR-AERO Triggering and Analysis application.
Group
Search and Mark
Syntax
SEARCH:SEARCH<x>:TRIGger:A :BUS:CAN:ERRType
{ACKMISS|BITSTUFFing|FORMERRor |ANYERRor}
SEARCH:SEARCH<x>:TRIGger:A :BUS:CAN:ERRType?
Arguments
ACKMISS species a search based on a missing ACK eld.
BITSTUFFing species a search based on a bit stufng error.
FORMERRor species a search based on a CAN FD form error. To use this option,
the CAN standard must be set to FDISO or FDNONISO .
ANYERRor species a search based on any error type.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-645

Table of Contents

Other manuals for Tektronix MSO54

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO54 and is the answer not in the manual?

Tektronix MSO54 Specifications

General IconGeneral
BrandTektronix
ModelMSO54
CategoryTest Equipment
LanguageEnglish

Related product manuals