Commands listed in alphabetical order
DISplay:PLOT
View<x>:CURSor:MODe
This command sets or queries the cursor tracking mode of the specified cursor
in the specified view.
Group
Cursor
Syntax
DISplay:PLOTView<x>:CURSor :MODe {INDEPENDENT|TRACK}
DISplay:PLOTView<x>:CURSor :MODe?
Arguments
INDEPENDENT allows independent adjustment of the two cursors.
TRACK ties the navigational functionality of the two cursors together. For cursor 1
adjustments, this ties the movement of the two cursors together; however, cursor 2
continues to move independently of cursor 1.
Examples
DISplay:PLOTView1:CURSor:M ODe TRACK sets the cu
rsor to track together.
DISplay:PLOTView1:CURSor:M ODe? might return
:DISPLAY:PLOTVIEW1:CURSOR: MODE INDEPENDENT
indicating independent
adjustment of the cursors is allowed.
DISplay:PLOTView<x>:CURSor:ONEOVERDELTATVALUE? (Query Only)
This command sets or queries the one over delta T cursor readout value of the
specified cursor in the specifi ed view.
Group
Cursor
Syntax
DISplay:PLOTView<x>:CURSor :ONEOVERDELTATVALUE?
Returns
Return an <NR3> that is the one over delta T cursor readout value (it may not be
delta T, depending o n the plot units).
Examples
DISplay:PLOTView<x>:CURSor :ONEOVERDELTATVALUE? might return
:DISPLAY:PLOTVIEW1:CURSOR: ONEOVERDELTATVALUE 88.229 5E+6
indicating the 1 over delta time value is 88.2295 MHz.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-253