Commands listed in alphabetical order
TRIGGER:A:BUS
:B1:FLEXRAY:CYCLECOUNT:VALUE ?
might return
TRIGGER:A:BUS:B1:FLEXRAY: CYCLECOUNT:VALUE "XXXXXX" indicating the
cycle count value is don't care.
TRIGger:{A|
B}:BUS:B<x>:FLEXray:DATa:HIVALue
This command specifies the high value when triggering on the FlexRay bus
data field. (Us e TRIGger:{A|B}:BUS:B<x>:FLEXray:DATa:VALue to set the
lower valu
e.) The trigger condition needs to be set to ID or IDANDDATA (using
TRIGger:{A|B}:BUS:B<x>:FLEXray:CONDition).The bus number is specified
by <x>.
Conditions
Requires the SR-AUTO Triggering and Analysis application.
Group
Trigger
Syntax
TRIGger:{A|B}:BUS:B<x>:FL EXray:DATa:HIVALue <QStri ng>
TRIGger:{A|B}:BUS:B<x>:FL EXray:DATa:HIVALue?
Arguments
<QString> is a quoted string that is the binary data high value.
Examples
TRIGGER:A:BUS:B1:FLEXRAY: DATA:HIVALUE
"11001101XXXXXXXXXXXXXXXX XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
XXXXXXX1"
sets the binary data string high value to
"11001101XXXXXXXXXXXXXXXX XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
XXXXXXX1"
.
TRIGGER:A:BUS:B1:FLEXRAY: DATA:HIVALUE? might
return
TRIGGER:A:BUS:B1:FLEXRAY:DA TA:HIVALUE
"XXXXXXXXXXXXXXXXXXXXXXXX XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
XXXXXXXX"
indicating the binary data string high value is don't care.
TRIGger:{A|B}:BUS:B<x>:FLEXray:DATa:OFFSet
This command specifies the offset of the data string, in bytes, when triggering
on the FlexRay bus data field. The trigger condition needs to be set to ID or
IDANDDATA (using TRIGger:{A|B}:BUS:B<x>:FLEXray:CONDition). The
bus number is specified by <x>.
Conditions
Requires the SR-AUTO Triggering and Analysis application.
2-796 MSO54, MSO56, MSO58, MSO58LP Programmer