Commands listed in alphabetical order
MEASUrement:
MINUI
This command sets or queries the minimum number of unit intervals required
for BUJ analysis.
Group
Measurement
Syntax
MEASUrement:MINUI <NR3>
MEASUrement:MINUI?
Arguments
<NR3> is the minimum number o f unit intervals required for BUJ analysis.
Examples
MEASUREMENT:MINUI 1.0000E+6 sets the minimum number of unit intervals
required for BUJ analysis to 1,000,000.
MEASUREMENT:MINUI? might return :MEASU REMENT:MINUI 1.0000E+6
indicating the minimum number of unit intervals required for BUJ analysis is
1,000,000.
MEASUrement:POPUlation:LIMIT:STATE
This command sets or queries the global population limit state for the
measurement.
Group
Measurement
Syntax
MEASUrement:POPUlation:LIM IT:STATE {OFF|ON|0|1 }
MEASUrement:POPUlation:LIM IT:STATE?
Arguments
OFF turns o ff the population limit.
ON turns on the population limit.
0 turns off the population limit.
1 turns on the population limit.
Examples
MEASUREMENT:POPULATION:LIM IT:STATE 1 turns on the popu
lation limit.
MEASUREMENT:POPULATION:LIM IT:STATE? might return
:MEASUREMENT:POPULATION:LI MIT:STATE 0 indicating the
population limit is off.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-495