Commands listed in alphabetical order
Syntax
MEASUrement:M
EAS<x>:RESUlts:CURRentacq:ST DDev?
Returns
The standard deviation of values accumulated for the specified measurement
since the last statistical reset.
Examples
MEASUrement:MEAS<x>:RESUlt s:CURRentacq:STDDev? might return
:MEASUREMENT:MEAS1:RESULTS :CURRENTACQ:STDDEV 0.0E+0 indicating
the standard devi
ation for the measurement is 0.0 V.
MEASUrement:MEAS<x>:SIGNALType
This command sets or queries the signal type of source 1 for the measurement.
The measure ment number is specified by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:SIGNAL Type {CLOCK|DATA|AUTO}
MEASUrement:MEAS<x>:SIGNAL Type?
Arguments
CLOCK specifies a clock signal type.
DATA specifies a data signal type.
AUTO automatically selects the signal type.
Examples
MEASUrement:MEAS1:SIGNALTy pe CLOCK specifies a clock signal type.
MEASUrement:MEAS1:SIGNALTy pe? might return
:MEASUREMENT:MEAS1:SIGNALT YPE AUTO indicating the instrument
automatically selects the signal type.
MEASUrement:MEAS<x>:SOUrce<x>
This command sets or queries the measurement source. The measurement number
and s ource a re specified by x.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:SOUrce <x>
MSO54, MSO56, MSO58, MSO58LP Programmer 2-485