Commands listed in alphabetical order
Examples
BUS:B1:SPI:DA
Ta:THReshold 50.0e-3
sets the threshold to 50.0 mV.
BUS:B1:SPI:DATa:THReshold ? might return
:BUS:B1:SPI:DATA:THRESHOL D 0.0E+0 indicating the threshold is set to
0.0 V.
BUS:B<x>:SPI:FRAMING
This command sets or queries the SPI framing setting for the specified bus. The
bus number is specified by x.
Conditions
Requires option SR-EMBD.
Group
Bus
Syntax
BUS:B<x>:SPI:FRAMING {IDLE|SS}
BUS:B<x>:SPI:FRAMING?
Arguments
IDLE specifies IDLE SPI framing.
SS specifies SS SPI framing.
Examples
BUS:B1:SPI:FRAMING IDLE sets the SPI framing to IDLE.
BUS:B1:SPI:FRAMING? might return :BUS:B1:SPI:FRAMING S S, indicating
that the SPI framing is set to SS.
BUS:B<x>:SPI:IDLETime
This command sets or queries the SPI idle time for the specified bus. The bus
is specified by x.
Conditions
Requires option SR-EMBD.
Group
Bus
Syntax
BUS:B<x>:SPI:IDLETime <NR3>
BUS:B<x>:SPI:IDLETime?
Arguments
<NR3> specifies the SPI idle time.
2-174 MSO54, MSO56, MSO58, MSO58LP Programmer