Commands listed in alphabetical order
MEASUREMENT:R
EFLEVELS:TYPE
? might return
:MEASUREMENT:REFLEVELS:TYP E GLOBAL indicating refer ence levels are
shared across measurements.
MEASUrement
:REF<x>:REFLevels:ABSolute:FALLHigh
This command sets or queries the value used as the high reference level of the
falling edge when the measurement's ref level method is set to absolute.
Group
Measurement
Syntax
MEASUrement:REF<x>:REFLeve ls:ABSolute:FALLHigh <NR3>
MEASUrement:REF<x>:REFLeve ls:ABSolute:FALLHigh?
Arguments
<NR3> is the value used as the h igh reference level of the falling edge when the
measurement's ref level method is set to absolute.
Examples
MEASUrement:REF1:REFLevels :ABSolute:FALLHigh 1.5 sets the
reference level to 1.5 V.
MEASUrement:REF1:REFLevels :ABSolute:FALLHigh? might return
:MEASUREMENT:REF1:REFLEVEL S:ABSOLUTE:FALLHIGH 1.000 0 indicating
the reference level is 1.0 V.
MEASUrement:REF<x>:REFLevels:ABSolute:FALLLow
This command sets or que ries the value used as the low reference l evel of the
falling edge when the measurement's ref level method is set to absolute.
Group
Measurement
Syntax
MEASUrement:REF<x>:REFLeve ls:ABSolute:FALLLow <NR3>
MEASUrement:REF<x>:REFLeve ls:ABSolute:FALLLow?
Arguments
<NR3> is the value used as the low reference level of the falling e dge when the
measurement's ref level method is set to absolute.
Examples
MEASUrement:REF1:REFLevels :ABSolute:FALLLow -1.5 sets the low
reference level to -1.5 V.