Commands listed in alphabetical order
BUS:B<x>:PAR
allel:ALLTHResholds:APPly (No Query Form)
This command sets all of the data source thresholds to the value set by
BUS:B<x>:PARallel:ALLTHResholds for the parallel bus. The bus is specied
by x.
Group
Bus
Syntax
BUS:B<x>:PARallel:ALLTHRe sholds:APPly
Related Commands
BUS:B<x>:PARallel:ALLTHResholds
Examples
BUS:B1:PARallel:ALLTHResh olds:APPly sets all data source thresholds to
the value set by BUS:B<x>:PARallel:ALLTHResholds.
BUS:B<x>:PARallel:BIT<n>SOUrce
This command sets or queries the specied bit source for specied parallel bus.
The bus is specied by x. The bit is specied by n and is an integer in the range of
1to64.
Group
Bus
Syntax
BUS:B<x>:PARallel:BIT<n>S OUrce
{CH<x>|CH<x>_D<x>|MATH<x> |REF<x>|REF<x>_D<x>|NONE}
Related Commands
BUS:B<x>:PARallel:BIT<n>SOUrce:THReshold
Arguments
CH<x> is the specied bit source.
CH<x>_D<x> is the specied bit source.
MATH<x> is the specied bit source.
REF<x> is the specied bit source.
REF<x>_D<x> species a digital reference waveform as the bit<x> source
waveform for the specied parallel bus.
NONE disables the bit source.
2-160 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Other manuals for Tektronix MSO54

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO54 and is the answer not in the manual?

Tektronix MSO54 Specifications

General IconGeneral
BrandTektronix
ModelMSO54
CategoryTest Equipment
LanguageEnglish

Related product manuals