Commands listed in alphabetical order
Syntax
MEASUrement:C
H<x>:REFLevels:ABSolute:FALL Low <NR3>
MEASUrement:CH<x>:REFLevel s:ABSolute:FALLLow?
Arguments
<NR3> is the high reference level, and is the zero percent level when
MEASUrement:IMMed:REFLevel:METHod is set to Absolute.
Examples
MEASUREMENT:CH2:REFLEVELS: ABSOLUTE:FALLLOW 1.5 sets the low
reference level of the falling e dge to 1.5 V.
MEASUREMENT:CH2:REFLEVELS: ABSOLUTE:FALLLOW? might return
:MEASUREMENT:CH2:REFLEVELS :ABSOLUTE:FALLLOW 1.5000 indicating
that the low reference level of the falling edge is 1.5 V.
MEASUrement:CH<x>:REFLevels:ABSolute:FALLMid
This command sets or queries the value used as the mid reference level of the
falling edge when the source ref level method is set to absolute. The channel
number is specified by x.
Group
Measurement
Syntax
MEASUrement:CH<x>:REFLevel s:ABSolute:FALLMid <NR3>
MEASUrement:CH<x>:REFLevel s:ABSolute:FALLMid?
Arguments
<NR3> is the mid reference level used to calculate the mid reference level when
the measurement's Ref level method is set to Absolute.
Examples
MEASUREMENT:CH2:REFLEVELS: ABSOLUTE:FALLMID 0 sets the mid reference
level of the falling edge to 0.0 V.
MEASUREMENT:CH2:REFLEVELS: ABSOLUTE:FALLMID? might return
:MEASUREMENT:CH2:REFLEVELS :ABSOLUTE:FALLMID 0.0E+0 indicating
that the mid reference level of the falling edge is 0.0 V.
MEASUrement:CH<x>:REFLevels:ABSolute:HYSTeresis
This command sets or queries the value of the hysteresis of the reference level
when the sourc e ref level method is set to absolute. The channel number is
specified by x.
Group
Measurement
MSO54, MSO56, MSO58, MSO58LP Programmer 2-383