Commands listed in alphabetical order
TRIGger:A:TIM
EOut:LOGICQUALification?
might include
:TRIGGER:A:TIMEOUT:LOGICQ UALIFICATION OFF indicating that logic
qualification is off.
TRIGger:{A|
B}:TIMEOut:POLarity
When triggering using the TIMEOut trigger type, this commands specifies the
polarity to be used.
Group
Trigger
Syntax
TRIGger:{A|B}:TIMEOut:POL arity {STAYSHigh|STAYSLow |EITher}
TRIGger:{A|B}:TIMEOut:POL arity?
Related Commands
TRIGger:{A|B}:TIMEOut:SOUrce
TRIGger:{A|B}:TIMEOut:TIMe
Arguments
STAYSHigh — Trigger when the signal stays high during the timeout time
specified by the command TRIGger:{A|B}:TIMEOut:TIMe.
STAYSLow — Trigger when the signal stays low during the timeout time
specified by the command TRIGger:{A|B}:TIMEOut:TIMe.
EITher — Trigger when the signal is either high or low during the timeout time
specified by the command TRIGger:{A|B}:TIMEOut:TIMe.
Examples
TRIGger:A:TIMEOut:POLarit y STAYSHigh specifies to trigger when the
signal stays high during the timeout time when triggering using the TIMEOut
trigger type.
TRIGger:A:TIMEOut:POLarit y? might return EITHER, indicating that the
signal stays either high or low during the timeout time.
TRIGger:{A|B}:TIMEOut:SOUrce
When triggering using the TIMEOut trigger type, this command specifies the
source. The available sources are live channels and digital channels. The default
is channel 1. The timeout trigger type is selected using TRIGger:{A|B}:TYPe
TIMEOut.
Group
Trigger
2-866 MSO54, MSO56, MSO58, MSO58LP Programmer