Commands listed in alphabetical order
TRIGger:A:TIM
EOut:LOGICQUALification?
might include
:TRIGGER:A:TIMEOUT:LOGICQ UALIFICATION OFF indicating that logic
qualication is off.
TRIGger:{A|
B}:TIMEOut:POLarity
When triggering using the TIMEOut trigger type, this commands species the
polarity to be used.
Group
Trigger
Syntax
TRIGger:{A|B}:TIMEOut:POL arity {STAYSHigh|STAYSLow |EITher}
TRIGger:{A|B}:TIMEOut:POL arity?
Related Commands
TRIGger:{A|B}:TIMEOut:SOUrce
TRIGger:{A|B}:TIMEOut:TIMe
Arguments
STAYSHigh Trigger when the signal stays high during the timeout time
specied by the command TRIGger:{A|B}:TIMEOut:TIMe.
STAYSLow Trigger when the signal stays low during the timeout time
specied by the command TRIGger:{A|B}:TIMEOut:TIMe.
EITher Trigger when the signal is either high or low during the timeout time
specied by the command TRIGger:{A|B}:TIMEOut:TIMe.
Examples
TRIGger:A:TIMEOut:POLarit y STAYSHigh species to trigger when the
signal stays high during the timeout time when triggering using the TIMEOut
trigger type.
TRIGger:A:TIMEOut:POLarit y? might return EITHER, indicating that the
signal stays either high or low during the timeout time.
TRIGger:{A|B}:TIMEOut:SOUrce
When triggering using the TIMEOut trigger type, this command species the
source. The available sources are live channels and digital channels. The default
is channel 1. The timeout trigger type is selected using TRIGger:{A|B}:TYPe
TIMEOut.
Group
Trigger
2-866 MSO54, MSO56, MSO58, MSO58LP Programmer

Table of Contents

Other manuals for Tektronix MSO54

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO54 and is the answer not in the manual?

Tektronix MSO54 Specifications

General IconGeneral
BrandTektronix
ModelMSO54
CategoryTest Equipment
LanguageEnglish

Related product manuals