Commands listed in alphabetical order
HORizontal:M
AIn:INTERPRatio? (Query Only)
This query-only command returns the Horizontal interpolation ratio.
Group
Horizontal
Syntax
HORizontal:MAIn:INTERPRati o?
Examples
HORIZONTAL:MAIN:INTERPRATI O? might return
:HORIZONTAL:MAIN:INTERPRAT IO 10.0.
HORizontal:MODE
This command set or queries the horizontal operating mode.
Group
Horizontal
Syntax
HORizontal:MODE {AUTO|MANual}
HORizontal:MODE?
Related Commands
HORizontal:MODE:RECOrdlength, HORizontal:MODE:SAMPLERate,
HORizontal:MODE:SCAle
Arguments
AUTO selects the automatic horizontal model. Auto mode automatically adjusts
the sample rate and record length to provide a high acquisition rate in Fast Acq or
signal fidelity in analysis. Record length is read only.
MANUAL selects the manual horizontal model. Manual mode lets you change
the sample rate, horizontal s cale, and record le ngth. These values interact. For
example, when you change record length then the horizontal scale also changes.
Examples
HORIZONTAL:MODE AUTO sets the horizontal mode to auto.
HORIZONTAL:MODE? might return :HORI ZONTAL:MODE MANUAL, indicating that
the horizontal mode is manual.
HORizontal:MODe:AUTOmatic:FASTAcq:RECOrdlength:MAXimum:VALue
Sets or queries the horizontal FastAcq maximum record length.
MSO54, MSO56, MSO58, MSO58LP Programmer 2-349