Commands listed in alphabetical order
Arguments
<NR1> is the cus
tom b it rate.
Examples
BUS:B1:RS232C:BITRATE:CUST OM 9500 sets the bit rate for the RS-232C bus
B1 to 9500 bits-per-second.
BUS:B1:RS232C:BITRATE:CUST OM? might return
:BUS:B1:RS232C:BITRATE:CUS TOM 9500, indicating that the bit rate for the
RS-232C bus B1 is set to 9500 bits-per-second.
BUS:B<x>:RS232C:DATABits
This command sets or queries the RS-232C data width for bus<x>, where the bus
number is specified by <x>.
Conditions
Requires option SR-COMP.
Group
Bus
Syntax
BUS:B<x>:RS232C:DATABits {7,8,9}
BUS:B<x>:RS232C:DATABits?
Arguments
<NR1> specifies the number of bits in the RS-232C data frame.
Examples
BUS:B1:RS232C:DATABITS 8 sets the data bits to 8 for the RS-232C bus B1.
BUS:B1:RS232C:DATABITS? might return :BUS:B1:RS23 2C:DATABITS 8,
indicating that the data bits is set to 8 for the R S-232C bus B1.
BUS:B<x>:RS232C:DELIMiter
This command sets or queries the RS-232C string delimiter on bus <x>, where
the bus number is specified by <x>. This command only applies when Packet
view is turned On.
Conditions
Requires option SR-COMP.
Group
Bus
MSO54, MSO56, MSO58, MSO58LP Programmer 2-165