Commands listed in alphabetical order
Examples
BUS:B1:PARall
el:BIT1SOUrce CH1
sets the bit 1 source to channel 1.
BUS:B1:PARallel:BIT1SOUrce ? might return
:BUS:B1:PARA
LLEL:BIT1SOURCE CH1_D0
indicating the bit1 source is
CH1_D0.
BUS:B<x>:PARallel:BIT<n>SOUrce:THReshold
This command sets or queries the specied bit source threshold for the specied
parallel bus. The bus is specied by x. The bit is specied by n a nd is an integer
in the range of 1 to 64.
Group
Bus
Syntax
BUS:B<x>:PARallel:BIT<n>SO Urce:THReshold <NR3>
Related Commands
BUS:B<x>:PARallel:BIT<n>SOUrce
Arguments
<NR3> is the specied bit source threshold for the specied parallel bus.
Examples
BUS:B3:PARALLEL:BIT2SOURCE :THRESHOLD 1.0 sets the threshold of bit
source 2 of parallel Bus 3 to 1.0 V.
BUS:B1:PARALLEL:BIT4SOURCE :THRESHOLD? might return
:BUS:B1:PARALLEL:BIT4SOURC E:THRESHOLD 250.00000E-3 indicates the
threshold of bit source 4 of parallel Bus 1 is currently set to 250.0 mV.
BUS:B<x>:PARallel:CLOCk:EDGE
This command sets or queries the clock edge for the parallel bus. The bus is
specied by x.
Group
Bus
Syntax
BUS:B<x>:PARallel:CLOCk:ED GE {FALLING|RISING|EITHER}
BUS:B<x>:PARallel:CLOCk:ED GE?
Related Commands
BUS:B<x>:PARallel:CLOCkSOUrce,
MSO54, MSO56, MSO58, MSO58LP Programmer 2-161

Table of Contents

Other manuals for Tektronix MSO54

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO54 and is the answer not in the manual?

Tektronix MSO54 Specifications

General IconGeneral
BrandTektronix
ModelMSO54
CategoryTest Equipment
LanguageEnglish

Related product manuals