Commands listed in alphabetical order
indicating tha
t the Start/End value for USB bus trigger on split token field search
1 is set to DONTCare.
SEARCH:SEARCH<x>:TRIGger:A:BUS:USB:TOKENType
This command sets or queries the token type used to search a USB bus signal.
Thesearchnumberisspecified by <x>.
Conditions
Requires the SR-USB2 Serial Triggering and Analysis Application.
Group
Search and Mark
Syntax
SEARCH:SEARCH<x>:TRIGger:A :BUS:USB:TOKENType
{ANY|IN|OUT|SETUP|SOF}
SEARCH:SEARCH<x>:TRIGger:A :BUS:USB:TOKENType?
Arguments
ANY specifies the token type as Any (XX01).
IN specifies the token type as IN (1001).
OUT specifies the token type as OUT (0001).
SETUP specifies the token type as SETUP (1101).
SOF specifies the token type as Start Of Frame (0101).
Examples
SEARCH:SEARCH1:TRIGGER:A:B US:USB:TOKENTYPE IN sets the token type
for USB bus trigger search 1 to IN.
SEARCH:SEARCH1:TRIGGER:A:B US:USB:TOKENTYPE? might return
:SEARCH:SEARCH1:TRIGGER:A: BUS:USB:TOKENTYPE SOF, indicating that
the token type for USB bus trigger search 1 is set to Start Of Frame.
SEARCH:SEARCH<x>:TRIGger:A:EDGE:SLOpe
This command sets or queries the slope for an edge trigger search to d
etermine
where to place a mark. The search number is specified by <x>.
Group
Search and Mark
Syntax
SEARCH:SEARCH<x>:TRIGger:A :EDGE:SLOpe {RISe|FALL|EIThe r}
SEARCH:SEARCH<x>:TRIGger:A :EDGE:SLOpe?
MSO54, MSO56, MSO58, MSO58LP Programmer 2-713