Commands listed in alphabetical order
Arguments
PCIExpress spe
cifies the PCIExpress dirac model.
FIBREchannel specifies the FIBREchannel dirac model.
Examples
MEASUREMENT:DIR
ACMODEL FIBRECHANNEL
sets the dirac model to
FIBRECHANNEL.
MEASUREMENT:DIRACMODEL? might return :MEASUREMENT:DIRACMODEL
PCIEXPRES
S indicating the dirac model is set to PCIECPRESS.
MEASUrement:DISPLAYUnits
This command sets or queries the display units used for jitter summary
measurements.
Group
Measurement
Syntax
MEASUrement:DISPLAYUnits {SEConds|UNIT INtervals}
Arguments
SEConds set the display units to seconds.
UNITINtervals set the display units to unit intervals.
Examples
MEASUREMENT:DISPLAYUNITS UNITINTERVALS sets the display units to unit
intervals.
MEASUREMENT:DISPLAYUNITS? might return :MEASUREMENT:DISPLAYUN ITS
SECONDS
indicating the display units are seconds.
MEASUrement:EDGE<x>
This command sets or queries the type of the edge for the measurement.
Group
Measurement
Syntax
MEASUrement:EDGE<x> {RISE|FALL|BOTH }
Arguments
FALL specifies the falling edge of the waveform.
RISE specifies the rising edge of the waveform.
BOTH specifies both a rising and falling edge of the waveform.