Page #102 background imageLoading...
Page #102 background image
9 - 6 BeneVision N22/N19 Patient Monitor Operator’s Manual
9.6.2.5 Setting the Reminder Tones
When the alarm volume is set to zero, or the alarm is reset or switched off, the monitor issues a periodical
reminder tone. The reminder tone is enabled by default.
To set the reminder tone, follow this procedure:
1. Select the Main Menu quick key turn to the third page from the System column select Maintenance
input the required password select Ok.
2. Select the Alarm tab Pause/Reset tab.
3. Set Alarm Reminder.
On: the monitor issues reminder tones at a designated interval.
Re-alarm: if the alarm condition persists the acknowledged alarms marked with “√” will be re-
generated after the designated reminder tone interval .
Off: the monitor does not issue reminder tones at a designated interval. The acknowledged alarms
marked with “√” will be silenced.
4. Set Reminder Interval.
9.6.2.6 Setting Alarm Volume Escalation
The monitor provides the function of automatically escalating the alarm tone volume. When this function is
enabled, if an alarm is not reset within the designated delay time after the alarm occurs, the volume of the alarm
tone increases automatically.
To set the alarm volume escalation, follow this procedure:
1. Select the Main Menu quick key turn to the third page from the System column select Maintenance
input the required password select Ok.
2. Select the Alarm tab Audio tab.
3. Set Auto Increase Volume.
2 Steps: if an alarm is not reset within the designated delay time after the alarm occurs, the alarm
volume automatically increases by two levels.
1 Step: if an alarm is not reset within the designated delay time after the alarm occurs, the alarm
volume automatically increases by one level.
Off: if an alarm is not reset within the designated delay time after the alarm occurs, the volume of the
alarm tone does not change.
4. Select Increase Volume Delay to set the delay time of alarm volume escalation.
NOTE
The alarm volume escalation function is not applied to the latched alarms.

Table of Contents

Other manuals for Mindray BeneVision N22

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N22 and is the answer not in the manual?

Mindray BeneVision N22 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N22
CategoryMonitor
LanguageEnglish

Related product manuals