BeneVision N22/N19 Patient Monitor Operator’s Manual 30 - 3
30.2 rSO2 Safety Information
CAUTION
• rSO
2
monitoring is intended for use in individuals greater than 2.5 kg at risk for reduced-flow or no-
flow ischemic states.
• Do not use the rSO
2
value as the sole basis for making decisions regarding diagnosis or therapy
because the rSO
2
values represent a small volume of tissue beneath the sensor and may not reflect
oxygenation disturbances that occur elsewhere.
• Use only recommended or provided accessories. Use any other sensor will compromise accuracy.
• Use of an electrosurgical instrument in the vicinity of the monitor may interfere with the signal and
result in inaccurate rSO
2
measurement.
NOTE
• Environments with excessive ambient light such as bright sunlight or strong operating room
lighting may require loosely covering the area of the sensor with an opaque drape.
30.3 rSO
2
Measurement Limitations
If present, the following may cause inaccurate readings:
â–  Cardiogreen, indigo carmine or other intravascular dyes
â–  Carboxyhemoglobin or other dyshemoglobins
â–  Hemoglobinopathies
â–  Conjugated hyperbilirubinemia (direct)
â–  Myoglobin (Mb) in muscle tissues
30.4 rSO
2
Display
Each rSO
2
numeric area displays one rSO
2
channel, which is shown as below:
(7)
(6) (5)
(1)
(2) (3)
(4)
(8)
(9)

Table of Contents

Other manuals for Mindray BeneVision N22

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N22 and is the answer not in the manual?

Mindray BeneVision N22 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N22
CategoryMonitor
LanguageEnglish

Related product manuals