10.7.5 Saving the Current ST as Baseline
ST analysis requires valid samples. The first baseline is automatically stored when valid ST values appear for 5
minutes. To update ST baselines, follow this procedure:
1. From the ST View window, select Set Baseline.
2. From the pop-up dialog box, select Ok to set the current ST segments and values as the baseline.
From the ST View window, you can also perform the following operations:
Display or hide ST baseline by selecting Display Baseline or Hide Baseline.
Display or hide the position of ISO point, J point and ST point by selecting Display Marker or Hide Marker.
CAUTION
Updating ST baseline affects ST alarms.
10.7.6 Displaying ST Segments in the Waveform Area
You can display ST segments in the waveform area. To do so, follow this procedure:
1. Access Tile Layout by either of the following ways:
Select the Screen Setup quick key select the Tile Layout tab.
Select Main Menu quick key from the Display column select Tile Layout.
2. Select the waveform area where you want to display the ST segments, and then select ST ST Segment.
10.7.7 Entering the ST Graphic Window
To display ST Graphic window, follow this procedure:
1. Select the ECG numeric area or waveform area to enter the ECG menu.
2. Select the ST tab.
3. From the bottom of the menu, select ST Graphic.