BeneVision N22/N19 Patient Monitor Operator’s Manual 10 - 19
10.7.5 Saving the Current ST as Baseline
ST analysis requires valid samples. The first baseline is automatically stored when valid ST values appear for 5
minutes. To update ST baselines, follow this procedure:
1. From the ST View window, select Set Baseline.
2. From the pop-up dialog box, select Ok to set the current ST segments and values as the baseline.
From the ST View window, you can also perform the following operations:
â– Display or hide ST baseline by selecting Display Baseline or Hide Baseline.
â– Display or hide the position of ISO point, J point and ST point by selecting Display Marker or Hide Marker.
• Updating ST baseline affects ST alarms.
10.7.6 Displaying ST Segments in the Waveform Area
You can display ST segments in the waveform area. To do so, follow this procedure:
1. Access Tile Layout by either of the following ways:
◆ Select the Screen Setup quick key → select the Tile Layout tab.
◆ Select Main Menu quick key → from the Display column select Tile Layout.
2. Select the waveform area where you want to display the ST segments, and then select ST→ ST Segment.
10.7.7 Entering the ST Graphic Window
To display ST Graphic window, follow this procedure:
1. Select the ECG numeric area or waveform area to enter the ECG menu.
2. Select the ST tab.
3. From the bottom of the menu, select ST Graphic.