BeneVision N22/N19 Patient Monitor Operator’s Manual 15 - 7
15.8.5 Enabling the NIBP End Tone
The monitor can issue a reminder tone at the completion of NIBP measurement. The NIBP End Tone is off by
default. To switch on the NIBP end tone, follow this procedure:
1. Select the NIBP numeric area to enter the
NIBP menu.
2. Select the
Setup tab.
3. Switch on
NIBP End Tone.
15.8.6 Setting NIBP Measurement Timeout
NIBP measurements become outline fonts after a preset time. This avoids older NIBP values being misinterpreted
as current measurements. To set the timeout period, follow this procedure:
1. Select the
Main Menu quick key → turn to the third page → from the System column select Maintenance
→ input the required password → select Ok.
2. Select the
Module tab → select Other.
3. Set
NIBP Timeout.
15.8.7 Displaying the NIBP List
To display multiple sets of the most recent NIBP measurements, follow this procedure:
1. Access
Tile Layout by either of the following ways:
◆ Select the Screen Setup quick key → select the Tile Layout tab.
◆ Select Main Menu quick key → from the Display column select Tile Layout.
2. In the desired numeric area, select
NIBP → NIBP List.
15.8.8 Correcting the NIBP Measurements
The middle of the cuff should be at the level of right atrium. If the limb is not at the heart level, you need to
correct the measurement:
â–  Add 0.75 mmHg (0.10 kPa) to the displayed value for each centimetre higher.
â–  Deduct 0.75 mmHg (0.10 kPa) to the displayed value for each centimeter lower.
15.9 Assisting Venous Puncture
You can use the NIBP cuff to cause sub-diastolic pressure to block the venous blood vessel and therefore help
venous puncture. To assist venous puncture, follow this procedure:
1. Select the
VeniPuncture quick key or select the NIBP numeric area → Setup tab.
2. Set
Venipuncture pressure.
3. Select
VeniPuncture at the bottom of the menu.
4. Puncture vein and draw blood sample.
5. Select the
NIBP Start/Stop quick key to deflate the cuff. If you do not deflate the cuff, the cuff automatically
deflates after a period of time (180 seconds for adult and pediatric patient, 90 seconds for neonatal patient).
During venous puncture, pay attention to the cuff pressure and the remaining time displayd in the NIBP numerics
area.

Table of Contents

Other manuals for Mindray BeneVision N22

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N22 and is the answer not in the manual?

Mindray BeneVision N22 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N22
CategoryMonitor
LanguageEnglish

Related product manuals