BeneVision N22/N19 Patient Monitor Operator’s Manual 3 - 3
3.3 Setting Up the Equipment
Observance of this manual is a prerequisite for proper product performance and correct operation. It ensures
patient and operator safety.
3.3.1 Connecting the AC Mains
The monitor is powered by AC power supply. Before connecting the equipment to the AC mains, check that the
voltage and frequency ratings of the power line are the same as those indicated besides the AC power input.
To use the AC power source, follow this procedure:
1. Connect the female end of the power cord with the AC power input.
2. Connect the male end of the power cord with a wall AC outlet.
3. Check that the AC indicator is on.
The AC indicator is off if the AC mains is not connected. When AC mains is connected, the AC indicator is
illuminated in green.
WARNING
• Always use the accompanying power cord delivered with the monitor.
• Before connecting the equipment to the AC mains, check that the voltage and frequency ratings of
the power line are the same as those indicated besides the AC power input.
• Use the cable retainer to secure the power cord to prevent it from falling off.
• Use the battery if the integrity of the protective earth conductor or the protective earthing system in
the installation is in doubt.
3.3.2 Connecting the Input Devices
Connect the mouse, keyboard, navigation knob, and barcode scanner if necessary.
3.3.3 Connecting the SMR
To connect the SMR, use the SMR cable (PN: 009-005121-00 or 009-005122-00) to connect the monitor
connector on the rear of SMR to the SMR connector on the main unit.
3.3.4 Connecting Modules to the SMR
To connect a module to the SMR, follow this procedure:
1. With the module properly oriented, align the module insertion guide slot with the SMR insertion guide.
Push the module into the SMR until you hear a click.
2. Push the lock at the bottom of the module inwards to lock the module.

Table of Contents

Other manuals for Mindray BeneVision N22

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N22 and is the answer not in the manual?

Mindray BeneVision N22 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N22
CategoryMonitor
LanguageEnglish

Related product manuals