D - 8 BeneVision N22/N19 Patient Monitor Operator’s Manual
D.2.12 CO
2
Technical Alarm Messages
Alarm message Default priority
Indication on
alarm reset
Cause and solution
CO2 Module High Temp Low C Ambient temperature is too high or there is a
module failure.
1. Lower the operating temperature.
2. Replug the module.
3. If the alarm persists, the CO
2
module may fail,
contact your service personnel.
CO2 Module Low Temp Low C Ambient temperature is too high or there is a
module failure.
1. Lower the operating temperature.
2. Replug the module.
3. If the alarm persists, the CO
2
module may fail,
contact your service personnel.
CO2 Zero Failed Low C For mainstream CO
2
module, check the
connections between the adapter and CO
2
transducer. Wait till the sensor’s temperature
becomes stabilized, and then perform a zero
calibration again.
For sidestream CO
2
module, replug the module. If
the alarm persists, contact your service
personnel.
CO2 No Watertrap Low B Check the watertrap connections.
CO2 High Airway Pressure Low C 1. Check the airway pressure settings of the
ventilator/anesthesia machine.
2. Disconnect the module from the ventilator/
anesthesia machine.
3. Replug the module.
4. If the alarm persists, contact your service
personnel.
CO2 Low Airway Pressure Low C 1. Check the airway pressure settings of the
ventilator/anesthesia machine.
2. Disconnect the module from the ventilator/
anesthesia machine.
3. Replug the module.
4. If the alarm persists, contact your service
personnel.
CO2 High Barometric Low C The ambient pressure exceeds the operating
pressure range or CO
2
module fails.
1. Make sure that the ambient pressure meets the
specifications, and check for sources that affect
the ambient pressure.
2. Replug the module. If the alarm persists,
contact your service personnel.
CO2 Low Barometric Low C The ambient pressure exceeds the operating
pressure range or CO
2
module fails.
1. Make sure that the ambient pressure meets the
specifications, and check for sources that affect
the ambient pressure.
2. Replug the module. If the alarm persists,
contact your service personnel.

Table of Contents

Other manuals for Mindray BeneVision N22

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N22 and is the answer not in the manual?

Mindray BeneVision N22 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N22
CategoryMonitor
LanguageEnglish

Related product manuals