9 - 18 BeneVision N22/N19 Patient Monitor Operator’s Manual
9.14 Testing Alarms
The monitor automatically performs a selftest at startup. Check that the alarm lamp illuminates, one after the
other, in red, yellow, and cyan, and that an alarm tone is heard. This indicates that the visible and audible alarm
indicators functions correctly.
To further test individual measurement alarms, perform measurements on yourself or using a simulator. Adjust
alarm limits and check that appropriate alarm behaviour is observed.
9.15 Actions When an Alarm Occurs
When an alarm occurs, observe the following steps and take proper actions:
1. Check the patient’s condition.
2. Confirm the alarming parameter or alarm category.
3. Identify the source of the alarm.
4. Take proper action to eliminate the alarm condition.
5. Make sure the alarm condition is corrected.
For more information, see D Alarm Messages.