BeneVision N22/N19 Patient Monitor Operator’s Manual A - 19
A.14.9 ScvO
2
/SvO
2
 Specifications
Operating mode Interfaces with Edwards Vigilance II or Vigileo monitor
Measured parameter Consistent with CCO-related parameters outputted by Vigilance II or Vigileo 
monitor 
Parameter alarm SvO
2
, ScvO
2
Signal Output
Standard Meets the requirements of IEC 60601-1 for short-circuit protection and 
leakage current
Output impedance ≤1000
Isolation voltage 1500 VAC
SpO
2
 Analog Signal Output
Output voltage 0 to 10V (0 to 100%)
Output voltage error ±5%
SvO
2
/ScvO
2
-related Parameters Outputted by Vigilance II Monitor
Name Measurement Range Resolution
SaO
2
40 to 100% 1
VO
2
0 to 999 ml/min 1
O
2
EI 0.0 to 99.9% 0.1
SNR -10 to 20 dB 0.1
DO
2
0 to 2000 ml/min 1
SvO
2
0 to 99% 1
ScvO
2
0 to 99% 1
SQI 1 to 4 1
SvO
2
/ScvO
2
-related Parameters Outputted by Vigileo Monitor
Name Measurement Range Resolution
SvO
2
0 to 99% 1
ScvO
2
0 to 99% 1
SQI 1 to 4 1
Alarm Limit Range(%) Step (%)
SvO
2
 /ScvO
2
High ScvO
2
<60%: (Low limit + 5%) to 60%
ScvO
2
≥60%: (Low limit + 1%) to 99%
ScvO2<60%: 5%
ScvO2≥60%: 1%
SvO
2
 /ScvO
2
 Low ScvO
2
<60%: 0 to (High limit - 5%)
ScvO
2
≥60%: 60% to (Low limit - 1%)