EasyManua.ls Logo

Mindray BeneVision N22 - Accessing the Hemodynamics Menu; Changing CCO Settings; Changing the CCO Alarm Settings; Changing the SVR Unit

Mindray BeneVision N22
472 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BeneVision N22/N19 Patient Monitor Operator’s Manual 19 - 7
Parity: None
Stop Bits: 1
Data Bits: 8
Flow Control: 2 seconds
5. Enter the Analog Input Port Setup menu of the Vigileo monitor, and set the CVP as follows:
Parameter: CVP
Voltage Range: 0-5 v
Full Scale Range: 100 mmHg (13.3 kPa)
Simulated High Value:100 mmHg (13.3 kPa)
Simulated Low Value: 0 mmHg (0.0 kPa)
For more information, See the Vigileo operator’s manual for the operation of the monitor.
WARNING
Calibrate the Vigileo monitor before monitoring. See the Vigileo operator’s manual for the
calibration instructions.
NOTE
For the Vigileo monitor, Flow Control must be set to 2 seconds.
19.6 Accessing the Hemodynamics Menu
To accessing the Hemodynamics menu, follow this procedure:
1. Select the
CCO numeric area to enter the CCOmenu.
2. Select the
Hemodynamics button. For information on hemodynamics, see 32 Hemodynamics.
19.7 Changing CCO Settings
19.7.1 Changing the CCO Alarm Settings
To change the CCO alarm settings, follow this procedure:
1. Select the CCO numeric area to enter the
CCO menu.
2. Select the
Alarm tab.
3. Set the alarm properties of CCO and CCI:
Switch on or switch off the alarms for CCO and CCI.
Adjust the alarm limits and alarm priority for CCO and CCI.
Switch on or switch off the alarm recording for CCO and CCI.
19.7.2 Changing the SVR Unit
To change the SVR unit, follow this procedure:
1. Select the CCO numeric area to enter the
CCO menu.
2. Select the
Setup tab.
3. Set
SVR Unit.

Table of Contents

Other manuals for Mindray BeneVision N22

Related product manuals