13 - 8 BeneVision N22/N19 Patient Monitor Operator’s Manual
If the SpO
2
value is effective, the monitor also adjusts the QRS tone (pitch tone) according to the SpO
2
value. For
information, see
13.6.8 Setting the SpO2 Tone Mode.
13.7.3 Setting the PR Source
Current pulse source is displayed in the PR numeric area. The PR from current pulse source has the following
characteristics:
PR is monitored as system pulse and generates alarms when you select PR as the active alarm source.
PR is stored in the monitor’s database and reviewed in the graphic/tabular trends; in trend graphs, as the PR
curve is in the same color with that of the PR source , it is unlikely to distinguish the PR source ;
PR is sent via the network to the CMS, if available.
To set which pulse rate as PR source, follow this procedure:
1. Select the SpO
2
numeric area or waveform area to enter the SpO2 menu.
2. Select the
PR Setup tab.
3. Set
PR Source.
The
PR Source menu displays the currently available PR sources from top to bottom by priority. When you select
Auto, the system will automatically select the first option as the PR source. If the current PR source is unavailable,
the system will automatically switch
PR Source to Auto. When you select IBP, the system will automatically
select the first pressure label as the PR source.
13.8 SpO
2
Troubleshooting
This section lists the problems that might occur. If you encounter the problems when using the equipment or
accessories, check the table below before requesting for services. If the problem persists, contact your service
personnel.
NOTE
For the physiological and technical alarm messages, see D Alarm Messages.
Problem Solution
Do not see SpO
2
numeric area or waveform area
on the main screen
1. Check that the SpO
2
is set to display in the Screen Setup menu.
For more information, see
3.11.2 Displaying Parameter Numerics
and Waveforms
.
2. Check that if the SpO
2
parameter switch is enabled. If not, enable
the SpO
2
measurement.For more information, see 3.11.1 Switching
On or Off a Parameter
.
3. Check that the cable connections of SpO
2
sensor and the
extension cable are tight. Replace the SpO
2
sensor or the extension
cable if needed.
Dashes “- -” display in place of numerics. 1. Check that the cable connections of SpO
2
sensor and the
extension cable are tight. Replace the SpO
2
sensor or the extension
cable if needed.
2. Reconnect the SpO
2
sensor if the alarm SpO2 Sensor Off appears.
3. Check the PI value. If the PI value is too low, adjust the SpO
2
sensor,
or apply the sensor to the site with better perfusion.
4. Move the sensor to the place with weaker light, or cover the sensor
with shade cloth if the alarm
SpO2 Sensor Off appears.
Low amplitude SpO
2
signal 1. The SpO
2
sensor and NIBP cuff are placed on the same limb.
Change a monitoring site if necessary.
2. Check the PI value. If the PI value is too low. Adjust the SpO
2
sensor,
or apply the sensor to the site with better perfusion.
3. Check the sensor and its application site.

Table of Contents

Other manuals for Mindray BeneVision N22

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N22 and is the answer not in the manual?

Mindray BeneVision N22 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N22
CategoryMonitor
LanguageEnglish

Related product manuals