D - 14 BeneVision N22/N19 Patient Monitor Operator’s Manual
D.2.23 Technical Alarm Messages Related to Networked Monitoring
Note: XX refers to the department name, YY refers to the room number, and ZZ refers to the bed number.
D.2.24 Technical Alarm Messages Related to Telemetry Monitors and BP10 Modules
Alarm message Default priority
Indication on
alarm reset
Cause and solution
No CMS Low B The monitor is disconnected from the CMS.
Check the network connection.
View Bed XX YY-ZZ,
Network Disconnected.
Low A The network is interrupted when the monitor is
viewing the remote device. Check the network
connection.
Viewed by Bed XX YY-ZZ,
Network Disconnected.
Low A The network is interrupted when the monitor is
viewed by another remote device. Check the
network connection.
WLAN IP Address Conflict Low C Wireless network IP network conflicts. Check the
network settings.
LAN1 IP Address Conflict Low C Wired network LAN1 IP network conflicts. Check
the network settings.
Fail To Get WLAN IP
Address
Low C Unable to automatically obtain the wireless
network IP address. Check the network settings.
Fail To Get LAN1 IP
Address
Low C Unable to automatically obtain the wired
network LAN1 IP address. Check the network
settings.
Alarm message Default priority
Indication on
alarm reset
Cause and solution
Telemetry Disconnected High B The TM80 is power-off powered off, the monitor
or the TM80 is not connected to the Wi-Fi
network, or out of the network coverage.
1. Power on the TM80.
2. Connect the monitor and the TM80 to the
network.
3. Move the TM80 in the network coverage.
The exchanger network connected by the TM80
and the monitor does not support multicast data
transfer.
Contact your service personnel.
Telemetry NIBP
Disconnected
High B The BP10 is power- off, or out of the MPAN
coverage.
1. Power on the BP10.
2. Move the BP10 in the MPAN coverage area.
Telemetry Error High C An error occurred to the TM80.
Restart the TM80. If the problem still persists,
replace with a known good TM80.
Telemetry Low Battery Med C The TM80 battery charge is low.
Replace with a known good battery.
Telemetry Battery
Depleted
High C The TM80 battery charge is critically low.
Replace with known good batteries.
Telemetry Battery
Maintenance Required
Med C The TM80 lithium-ion battery may reach its life.
Replace with known good batteries.