BeneVision N22/N19 Patient Monitor Operator’s Manual D - 15
D.2.25 Other System Technical Alarm Messages
Telemetry Battery Error Med C The TM80 lithium-ion battery communication 
encounters an error. Replace with known good 
batteries.
Telemetry Battery Type 
Error
Med C The TM80 battery contacts may fail. Replace with 
known good batteries.
Telemetry NIBP Error High C An error occurred to the BP10. Restart the BP10. If 
the problem persists, replace with a known good 
BP10.
Telemetry NIBP Low 
Battery
Med C The BP10 battery charge is low. Replace with a 
known good battery.
Telemetry NIBP Battery 
Depleted
High C The BP10 battery charge is critically low. Replace 
with a known good battery.
Telemetry NIBP Battery 
Maintenance Required
Med C The BP10 lithium-ion battery is may reach its 
lifetime. Replace with a known good battery.
Telemetry NIBP Battery 
Error
Med C The BP10 lithium-ion battery communication 
encounters an error. Replace with a known good 
battery.
Telemetry NIBP Voltage 
Error
High C The BP10 battery voltage is abnormal. Replace 
with a known good battery.
Alarm message Default priority
Indication on 
alarm reset
Cause and solution
Alarm message Default priority
Indication on 
alarm reset
Cause and solution
Primary Screen Alarm 
Error
High C The alarm signal communication between the 
primary display and the main unit interrupts. The 
primary display cannot provide alarm indications. 
Check the connection between the main unit and 
the primary display.
Secondary Screen Alarm 
Error
High C The alarm signal communication between the 
secondary display and the main unit interrupts. 
The secondary display cannot provide alarm 
indications. Check the connection between the 
main unit and the secondary display.
XX: Disconnected
(XX refers to the name of 
the external device)
High B Corresponding external device is disconnected. 
Check the connection between the monitor and 
the external device.
Storage Card Error High C The storage card fails or files are damaged. 
Restart the monitor to format the storage card. If 
the alarm persists, contact your service 
personnel.
Loading Default Config 
Failed
Low A The default configuration is not correctly loaded. 
The monitor will restore to the factory default 
configuration for the current patient category.
XX Conflicts
(XX refers to the module 
label)
Prompt / The same type of corresponding module being 
used exceeds the supported number. Remove 
the conflict module. 
XXX Measurement has 
been closed
(XX refers to the module 
label)
Prompt / The parameter module is disabled. Switch on the 
module if you want to use it. For more 
information, see 
3.11.1 Switching On or Off a 
Parameter
.