BeneVision N22/N19 Patient Monitor Operator’s Manual D - 15
D.2.25 Other System Technical Alarm Messages
Telemetry Battery Error Med C The TM80 lithium-ion battery communication
encounters an error. Replace with known good
batteries.
Telemetry Battery Type
Error
Med C The TM80 battery contacts may fail. Replace with
known good batteries.
Telemetry NIBP Error High C An error occurred to the BP10. Restart the BP10. If
the problem persists, replace with a known good
BP10.
Telemetry NIBP Low
Battery
Med C The BP10 battery charge is low. Replace with a
known good battery.
Telemetry NIBP Battery
Depleted
High C The BP10 battery charge is critically low. Replace
with a known good battery.
Telemetry NIBP Battery
Maintenance Required
Med C The BP10 lithium-ion battery is may reach its
lifetime. Replace with a known good battery.
Telemetry NIBP Battery
Error
Med C The BP10 lithium-ion battery communication
encounters an error. Replace with a known good
battery.
Telemetry NIBP Voltage
Error
High C The BP10 battery voltage is abnormal. Replace
with a known good battery.
Alarm message Default priority
Indication on
alarm reset
Cause and solution
Alarm message Default priority
Indication on
alarm reset
Cause and solution
Primary Screen Alarm
Error
High C The alarm signal communication between the
primary display and the main unit interrupts. The
primary display cannot provide alarm indications.
Check the connection between the main unit and
the primary display.
Secondary Screen Alarm
Error
High C The alarm signal communication between the
secondary display and the main unit interrupts.
The secondary display cannot provide alarm
indications. Check the connection between the
main unit and the secondary display.
XX: Disconnected
(XX refers to the name of
the external device)
High B Corresponding external device is disconnected.
Check the connection between the monitor and
the external device.
Storage Card Error High C The storage card fails or files are damaged.
Restart the monitor to format the storage card. If
the alarm persists, contact your service
personnel.
Loading Default Config
Failed
Low A The default configuration is not correctly loaded.
The monitor will restore to the factory default
configuration for the current patient category.
XX Conflicts
(XX refers to the module
label)
Prompt / The same type of corresponding module being
used exceeds the supported number. Remove
the conflict module.
XXX Measurement has
been closed
(XX refers to the module
label)
Prompt / The parameter module is disabled. Switch on the
module if you want to use it. For more
information, see
3.11.1 Switching On or Off a
Parameter
.

Table of Contents

Other manuals for Mindray BeneVision N22

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N22 and is the answer not in the manual?

Mindray BeneVision N22 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N22
CategoryMonitor
LanguageEnglish

Related product manuals