BeneVision N22/N19 Patient Monitor Operator’s Manual A - 23
A.14.13 ICG Specifications
A.14.14 CO
2
Specifications
Technique Thoracic electrical bioimpedance (TEB)
Measurement range SV:
HR:
C.O.:
5 to 250 ml
44 to 200 bpm
1.0 to 15 L/min
Accuracy SV:
HR:
C.O.:
Not specified.
±2 bpm
Not specified
Alarm limit Range Step
C.I. High (low limit + 1.0) to 15.0 L/min/m
2
0.1 L/min/m2
C.I. Low 1.4 to (high limit - 1.0)L/min/m
2
TFC High (low limit + 1) to 125/k 1 /k
TFC Low 19 to (high limit - 1)/k
Measurement mode Sidestream, microstream, mainstream
Technique Infrared absorption
Apnea time 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
Alarm limit Range Step
EtCO
2
High (low limit + 2) to 99 mmHg 1 mmHg
EtCO
2
Low 1 to (high limit - 2)mmHg
FiCO
2
High 1 to 99 mmHg
EtO
2
High (low limit + 2%) to 100% 1%
EtO
2
Low 18% to (high limit - 2)%
FiO
2
High (low limit + 2%) to 100%
FiO
2
Low 18% to (high limit - 2)%
awRR High Adult, pediatric:
awRR≤20
awRR>20
Neonate:
awRR≤20
awRR>20
(low limit + 2) to 20
(low limit + 5) to 100
(low limit + 2) to 20
(low limit + 5) to 150
awRR≤20: 1
awRR>20: 5
awRR Low awRR≤20: 0 to (high limit - 2)
awRR>20: 0 to (high limit - 5)