BeneVision N22/N19 Patient Monitor Operator’s Manual C - 9
C.6 IBP Default Settings
Item Default Setting
IBP-S Alarm switch On
High limit
Art/pArt/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
Adult: 160 mmHg
Pediatric: 120 mmHg
Neonate: 90 mmHg
PA
Adult: 35 mmHg
Pediatric and neonate: 60 mmHg
Low limit Art/pArt/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
Adult: 90 mmHg
Pediatric: 70 mmHg
Neonate: 55 mmHg
PA
Adult: 10 mmHg
Pediatric and neonate: 24 mmHg
Priority Med
Print on Alarm Off
IBP-D Alarm switch On
High limit
Art/pArt/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
Adult: 90 mmHg
Pediatric: 70 mmHg
Neonate: 60 mmHg
PA
Adult: 16 mmHg
Pediatric and neonate: 4 mmHg
Low limit
Art/pArt/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
Adult: 50 mmHg
Pediatric: 40 mmHg
Neonate: 20 mmHg
PA
Adult: 0 mmHg
Pediatric and neonate: -4 mmHg
Priority Med
Print on Alarm Off