EasyManua.ls Logo

Mindray BeneVision N22 - 31.2.10 Common Operations

Mindray BeneVision N22
472 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BeneVision N22/N19 Patient Monitor Operator’s Manual 31 - 11
31.2.10 Common Operations
This section describes the common operations that are applicable for all the review pages.
31.2.10.1 Displaying Two Review Pages Side by Side
You can display two review pages side by side to compare trend data and learn your patients condition changes
from different perspectives.
To display two review pages side by side, follow this procedure:
1. Enter the desired review page by one of the following methods:
Select the Review quick key select the desired tab.
Select the Main Menu quick key from the Review column select the desired menu item.
2. Select (for landscape display) or (for portrait display) at the upper left corner of the desired
review page. A new review page will be displayed and put side by side with the desired review page.
You can browse trend data, view events, select other review tabs on the two review pages. If you refresh the trend
data by moving cursor or on one review page, the trend data on the other review page will be updated
accordingly.
31.2.10.2 Browsing Trend Data
On a review page, you can browse trend data and carry out related tasks by one of the following methods:
Select the following buttons.
Selecting or moves the cursor one page to the left or right to navigate through the trends
timeline
Selecting or moves the cursor one step to the left or right to navigate through the trends
timeline.
Selecting or scrolls one page up or down the review page to see measurement trends that do
not fit in the current view.
Selecting or scrolls one step up or down the review page to see measurement trends
that do not fit in the current view.
Move the cursor.
Move .
31.2.10.3 Viewing Events
You can view the following types of events:
Manually triggered events
Parameter-related operation events and alarm-related events, such as starting C.O. measurement
Operation events not related to parameters, such as system time change
Events are displayed in descending chronological order. The most recent event is displayed at the top. Events
displayed include the event trigger time, event name, and alarm priority indicator. The number of asterisk
symbols before and event matches the following alarm priority:
***: high priority alarm
**: medium priority alarm
*: low priority alarm