5 - 2 BeneVision N22/N19 Patient Monitor Operator’s Manual
5.1 Configuration Introduction
When performing continuous monitoring on a patient, the clinical professional often needs to adjust the
monitor’s settings according to the patient’s condition. The collection of all these settings is called a
configuration. The system configuration items can be classified as: parameter configuration items, conventional
configuration items, and user maintenance items. Allowing you to configure the monitor more efficiently, the
monitor provides different sets of configurations to accommodate the varying patient categories and
departments. You can change some settings from a certain set of configuration and then save the changed
configuration as a user configuration.
The default configurations provided for your monitor are department-oriented. You can choose either from:
â–  General
â–  OR
â–  ICU
â–  NICU
â–  CCU
Each department has three different sets of configurations tailored for adult, pediatric and neonatal patients.
WARNING
• The configuration management function is password protected. The configuration management
tasks must be performed by clinical professionals.
5.2 Changing the Department
If the current department configuration is not the one you want to view, you can change the department by
following this procedure:
1. Select the Main Menu quick key → turn to the second page → from the Configuration column select
Manage → input the required password → Select Ok.
2. Select Change Department.
3. Select a department.
4. Select Ok.
CAUTION
• Changing the department will delete all current user configurations.
5.3 Setting Default Configuration
The monitor will load the pre-set default configuration in the following cases:
â–  A patient is admitted.
â–  A patient is discharged.
â–  Patient data is cleared.
â–  Patient category is changed.
To set the default configuration, follow this procedure:
1. Select the Main Menu quick key → turn to the second page → from the Configuration column select
Manage → input the required password → Select Ok.
2. Select Select Default Config.

Table of Contents

Other manuals for Mindray BeneVision N22

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N22 and is the answer not in the manual?

Mindray BeneVision N22 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N22
CategoryMonitor
LanguageEnglish

Related product manuals