29 - 2 BeneVision N22/N19 Patient Monitor Operator’s Manual
29.1 NMT Introduction
This monitor can connect a Organon TOF-Watch SX monitor for NMT(neuromuscular transmission) monitoring.
This monitor can display, store and review measurements from TOF-Watch SX monitor, as well as present related
alarms. On this monitor, you can separately set the level of NMT related alarms and switch on or off alarm
recording; you can also view TOF-Watch SX monitor settings of alarm limits and alarm switch.
29.2 NMT Safety Information
WARNING
• TOF-Watch SX monitor is manufactured by Organon. This company provides the technology for
measuring NMT parameters. We only provide the connection between this monitor and TOF-Watch
SX monitor.
• If you have any doubts about the operation and maintenance of the TOF-Watch SX monitor, please
see TOF-Watch SX monitor operator’s manual or directly contact Organon.
• Fully observe TOF-Watch SX monitor operator’s manual to make settings and to connect the monitor
with a patient.
29.3 Connecting a TOF-Watch SX monitor
The TOF-Watch SX monitor connects with BeneLink module through an ID adapter, see the figure below.
To connect the TOF-Watch SX monitor, follow this procedure:
1. Insert a BeneLink module into the SMR
2. Connect the ID adapter that matches the TOF-Watch SX monitor to the BeneLink module with an RJ45
connecting cable.
3. Connect the ID adapter to the TOF-Watch SX interface with Mindray type C serial port adapting cable (PN:
009-001769-00).
4. Connect the TOF-Watch SX interface to the TOF-Watch SX monitor.
5. Put a label indicating device name to the RJ45 connecting cable at the end near the BeneLink module.
When the BeneLink module is connected to several external devices, you can easily recognize the devices
with these labels.
(1) BeneLink Module (2) Label
(3) RJ45 connecting cable (4) ID Adapter
(5) Type C serial port adapting cable & TCM interface cable
(6) TCM monitor
(6)
(4)
(3)
(2)
(1)
(5)

Table of Contents

Other manuals for Mindray BeneVision N22

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N22 and is the answer not in the manual?

Mindray BeneVision N22 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N22
CategoryMonitor
LanguageEnglish

Related product manuals