C - 18 BeneVision N22/N19 Patient Monitor Operator’s Manual
C.16 EEG Default Settings
C.17 BIS Default Settings
Vol View MVe, MVi, TVe, TVi
Item Default Setting
Item Default Setting
Scale 100 v
Speed 25mm/s
Low Filter 0.5 Hz
High Filter 30 H
Notch Filter On
EEG Expand View Default Setting
EEG EEG Channels All
Scale 100 v
Speed 25mm/s
Trends EEG Channels EEG2, EEG3, EEG4
Parameters SEF
Trend Length 60 min
DSA EEG Channels All
Parameters SEF
Trend Length 20 min
Power Scale 1 - 64 dB
CSA EEG Channels All
Parameters SEF
Trend Length 20 min
Power Scale 1 - 64 dB
CSA Clipping On
Item Default Setting
BIS Alarm switch On
High limit 70
Low limit 20
Priority Med
Print on Alarm Off
Smoothing Rate 15 sec
Display OR: BIS Trend
Other departments: EEG LT/RT
Scale 100 v
Speed 25mm/s