A - 20 BeneVision N22/N19 Patient Monitor Operator’s Manual
A.14.10 CCO Specifications
Operating mode Interfaces with Edwards Vigilance II or Vigileo monitor
Measured parameter Consistent with CCO-related parameters outputted by Vigilance II or Vigileo
monitor
Parameter alarm Vigilance II: CCO, CCI, EDV, EDVI,SVR, SVRI, SV, SVI, RVEF
Vigileo: CCO, CCI, SV, SVI, SVRI, SVV
Signal Outputs
Standard Meets the requirements of IEC 60601-1 for short-circuit protection and
leakage current
Output impedance ≤1000
Isolation voltage 1500 VAC
ECG Analog Output
Bandwidth (-3dB; reference frequency:
10Hz)
ST mode: 0.05~40Hz
Diagnostic mode: 0.05~150Hz
Monitor mode: 0.5~40Hz
Surgical mode: 1~20Hz
Sensitivity 2V/mV ±5%
MAP Analog Signal Output
Output voltage DC 0 to 5V (0 to 500mmHg)
Output voltage error ±5%
CVP Analog Signal Output
Output voltage DC 0 to 5V (0 to 100mmHg)
Output voltage error ±5%
CCO-related Parameters Outputted by Vigilance II Monitor
Name Range Resolution
CCO 1 to 20 L/min 0.1 L/min
CCI 0 to 20 L/min/m
2
0.1 L/min/m
2
CO 1 to 20 L/min 0.1 L/min
CI 0 to 20 L/min/m
2
0.1 L/min/m
2
EDV 40 to 800 ml 1 ml
EDVI 20 to 400 ml/m
2
1 ml/m
2
SVR 0 to 3000 DS/cm
5
0.1 kPa-s/l
SVRI 0 to 6000 DS•m
2
/cm
5
0.1kPa-m
2
/l
SV 0 to 300 ml 1 ml
SVI 0 to 200 ml/m
2
1 ml/m
2
TB 25 to 45 °C 0.1 °C, 0.1°F
RVEF 10 to 60% 1%
ESV 10 to 700 ml 1 ml
ESVI 5 to 400 ml/m
2
1 ml/m
2