BeneVision N22/N19 Patient Monitor Operator’s Manual 19 - 9
19.8 Changing SvO
2
/ScvO
2
Settings
19.8.1 Changing the SvO
2
/ScvO
2
Alarm Settings
To change the SvO
2
/ScvO
2
alarm settings, follow this procedure:
1. Select the SvO
2
/ScvO
2
numeric area to enter the SvO2 or ScvO2 menu.
2. Select the
Alarm tab.
3. Set the alarm properties of SvO
2
/ScvO
2
:
â—† Switch on or switch off the alarms for SvO
2
/ScvO
2
.
â—† Adjust the alarm limits and alarm priority for SvO
2
/ScvO
2
.
â—† Switch on or switch off the alarm recording for SvO
2
/ScvO
2
.
19.8.2 Setting the SvO
2
/ScvO
2
Analog Output Signal
This monitor can output the analog signals to the Vigilance II monitor. To set the SvO
2
/ScvO
2
analog output
signal, follow this procedure:
1. Select the SvO
2
numeric area to enter the SvO2 menu; or select the ScvO
2
numeric area to enter the ScvO2
menu.
2. Select the
Signal Output Setup tab.
3. Set the output signal as follows:
â—† This monitor can output the analog signals of ECG waveform, MAP value, SpO
2
value and CVP value to
the Vigilance II monitor. If a signal has several sources, you can select a source.
â—† Select Simulated High Value to output the simulated high value calibration signals to Vigilance II
monitor. To stop output the simulated high value signal, select
Simulated High Value again.
â—† Select Simulated Low Value to output the simulated low value calibration signals to Vigilance II
monitor. To stop output the simulated low value signal, select
Simulated Low Value again.
The following table shows values and voltages of the high and low value calibration signals.
NOTE
• See the operator’s manual of Vigilance II monitor for the calibration instructions.
Parameters Parameter Values Output Voltage
High Value Calibration Signals
MAP 500 mmHg 5V
SpO
2
100% 10V
CVP 100mmHg 5V
Low Value Calibration Signals
MAP 0 mmHg 0V
SpO
2
0% 0V
CVP 0mmHg 0V

Table of Contents

Other manuals for Mindray BeneVision N22

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N22 and is the answer not in the manual?

Mindray BeneVision N22 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N22
CategoryMonitor
LanguageEnglish

Related product manuals