3 - 10 BeneVision N22/N19 Patient Monitor Operator’s Manual
.
• In Privacy mode, all audible alarms are suppressed and the alarm light is deactivated at the monitor.
Alarms are presented only at the CMS. Pay attention to potential risk.
• The privacy mode is not available if the Department is set to OR.
• You cannot enter the privacy mode if a low battery alarm occurs.
3.7.2.2 Exiting the Privacy Mode
The monitor automatically exit the privacy mode in any of the following situations:
â– The monitor disconnects from the CMS.
â– The low battery alarm occurs.
You can also operate the touchscreen, mouse, or keyboard to manually exit the privacy mode.
3.7.3 Night Mode
The night mode is a special clinical mode. To avoid disturbing the patient, you can use the night mode.
3.7.3.1 Entering the Night Mode
To enter the night mode, follow this procedure:
1. Select the Night Mode quick key, or select the Main Menu quick key → from the Display column select
Night Mode.
2. Change the night mode settings if necessary.
3. Select Enter Night Mode.
The night mode settings are as follows by default:
â– Brightness: 1
â– Alarm Volume: 2
â– QRS Volume: 1
â– Key Volume: 0
â– NIBP End Tone: Off
â– Stop NIBP: Off
• Verify the night mode settings before entering the night mode. Pay attention to the potential risk if
the setting value is low.