D - 12 BeneVision N22/N19 Patient Monitor Operator’s Manual
D.2.19 Battery
Technical Alarm Messages
D.2.20 Recorder Technical Alarm Messages
rSO2: Change Sensor Low C The sensor type mismatches patient category,
the reusable sensor cable fails, the sensor reaches
the lifetime, or the sensor type mismatches the
monitor and patient category setting. Verify
patient category and use sensors of correct
patient category.
rSO2 -1/rSO2-2
Interference
Low C There is noise interference (such as
electrosurgical unit). Check for any possible
sources of signal noise.
rSO2-1 Auto Baseline Prompt / The monitor automatically sets the rSO
2
baseline.
rSO2-2 Auto Baseline
Disconnect/ Reconnect
rSO2
Low A Remove the rSO
2
module and reconnect it.
Alarm message Default priority
Indication on
alarm reset
Cause and solution
Alarm message Default priority
Indication on
alarm reset
Cause and solution
Low Battery Med C Connect the monitor to an AC power source and
allow the batteries to charge.
Critically Low Battery High C Connect the monitor to an AC power source and
allow the batteries to charge.
Battery Service Required Low B The battery reaches its lifetime. Replace the
battery.
Battery Overload High C Too many parameter modules are connected,
causing system overload and high power
consumption. Use AC power supply.
Power Board Comm Error High C Restart the monitor. If the alarm persists, contact
your service personnel.
Battery Error High C Replace the battery.
Alarm message Default priority
Indication on
alarm reset
Cause and solution
Recorder Init Error Low A An error occurred during the recorder
initialization. Replug the module. If the alarm
persists, contact your service personnel.
Recorder Comm Error Low A Replug the recorder and restart the monitor if not
solved. If the alarm persists, contact your service
personnel.
Recorder Unavailable Low A Recorder module failure. Replace the module.
Recorder Head Hot Low C The recorder has been working for too long time.
Stop the recording and resume the recording till
the recorder’s print head cools down.
Recorder Initializing Prompt / Wait until the recorder initialization is completed.
Recorder out of Paper Prompt / The recorder paper is not loaded or the recorder
door is not closed. Check the recorder, load the
recorder paper or close the recorder door.