BeneVision N22/N19 Patient Monitor Operator’s Manual 15 - 3
• Continuous cuff pressure due to connection tubing kinking may cause blood flow interference, and
resulting in harmful injury to the patient.
• NIBP reading can be affected by the measurement site, the position of the patient, exercise, or the
patient's physiologic condition. If you doubt the NIBP measurements, determine the patient’s vital
signs by alternative means, and then verify that the monitor is working correctly.
• Devices that exert pressure on tissue have been associated with purpura, ischemia, and neuropathy.
Inspect the application site regularly to ensure skin quality and inspect the extremity of the cuffed
limb for normal color, warmth and sensitivity. If the skin quality changes, or if the extremity
circulation is being affected, move the cuff to another site or stop the blood pressure measurements
immediately. Check more frequently when making automatic or STAT measurements. Auto NIBP
measurements with one and two minute intervals are not recommended for extended periods of
time.
• NIBP diagnostic significance must be decided by the physician.
CAUTION
• Only use parts and accessories specified in this manual. Follow the instructions for use and adhere to
all warnings and cautions.
• Accuracy of NIBP measurement depends on using a cuff of proper size. It is essential to measure limb
circumference and choose a cuff with proper size.
15.3 NIBP Measurement Limitations
Measurements are impossible with heart rate extremes of less than 30 bpm or greater than 300 bpm, or if the
patient is on a heart-lung machine. The measurement may be inaccurate or impossible in the following
situations:
â–  Regular arterial pressure pulses are hard to detect
â–  With excessive and continuous patient movement such as shivering or convulsions
â–  With cardiac arrhythmias
â–  With rapid blood pressure changes
â–  With severe shock or hypothermia that reduces blood flow to the peripheries
â–  On an edematous extremity.
NOTE
• The effectiveness of this sphygmomanometer has not been established in pregnant, including pre-
eclamptic patients.
15.4 Measurement Modes
There are three NIBP measurement modes:
â–  Manual: measurement on demand.
â–  Auto: repeated measurements at set interval.
â–  STAT: continually rapid series of measurements over a five minute period.

Table of Contents

Other manuals for Mindray BeneVision N22

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N22 and is the answer not in the manual?

Mindray BeneVision N22 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N22
CategoryMonitor
LanguageEnglish

Related product manuals